Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-00865-3 |
Product Name/Description |
Canon CXDA - 70C Wireless Computed radiographic system
ARTG Number: 176946 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
27/08/2012 |
Responsible Entity |
|
Reason/Issue |
It has been reported that in certain cases screws securing the electronic circuit board inside the product became loose and fell out, potentially damaging the internal electronic circuit board, resulting in malfunctions including a loss of power in some cases and possible image failures.
This recall action was not notified to the TGA before it was initiated by Canon Australia. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Canon Australia is initiating a recall for product correction to replace all affected product sensor units. |
Contact Information |
02 8873 8611 - Canon Australia |