Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-00824-3 |
Product Name/Description |
Halo Flex Energy Generators (Endotherapy electrosurgical unit)
Model Number: 1190A-230A All serial numbers ARTG Number: 140684 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
14/08/2012 |
Responsible Entity |
|
Reason/Issue |
Covidien has received reports related to a generator malfunction where the HALOFLEX Energy Generator fails to enter “standby mode” when initially powered-on. The generator became stuck in the self-test mode upon initial power-on, did not enter standby mode, and therefore could not be used for patient care. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
The customer letter provides temporary work around instructions. Covidien is developing a software revision, which will be available by the end of August 2012 that will prevent the malfunction from recurring. |
Contact Information |
1800 350 702 - Technical Services |