Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-00805-3 |
Product Name/Description |
Dash 3000/4000/5000 Patient Monitor with software version 6.0 and later (multiparameter patient monitor).
Monitors with software versions 6.0 and later may be affected
ARTG Number: 92726 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
8/08/2012 |
Responsible Entity |
|
Reason/Issue |
The Dash monitor may exhibit an unexpected loss of Custom Automatic NBP Measurement cycling. Custom Automatic NBP Measurement is a feature introduced in Dash version 6.0 which allows the user to program up to 4 different NBP auto mode sequences. When a view of another patient with an NBP parameter on the network occurs, if the Custom Automatic NBP Measurement cycling function is enabled, the function will turn off unexpectedly. The issue can occur when manually viewing another bed on the network using the View Other Beds option or automatically with the Automatic View On Alarm (AVOA) feature. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
GE Healthcare is providing users with work around instructions to mitigate the issue. A software update will be implemented to permanently fix the issue when available. |
Contact Information |
1300 722 229 - GE Healthcare National Service and Support Centre |