Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-00706-3 |
Product Name/Description |
Post Fuse Assembly with Power Wheelchairs (electronic wheelchair)
Multiple Model Numbers. (Distributed between 11/10/2010 and 21/10/2010)
ARTG 160385 |
Recall Action Level |
Consumer |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
13/08/2012 |
Responsible Entity |
|
Reason/Issue |
Post fuse assemblies on certain models of Power Wheelchairs could potentially be defective in a manner that would result in the fuse not operating as intended (blowing) in the event of future damage or faults within the main wiring harness. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Distributors to rework affected chairs with the retrofit kits provided by the Sponsor. |
Contact Information |
1800 460 460 - Invacare Australia Customer Service |