Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-00699-3 |
Product Name/Description |
Dimension RxL Reagent Management System (RMS) Power Cord Plug; An in vitro diagnostic medical device (IVD)
Model Numbers: 10444832, 10444833 and 10453482
Manufactured prior to April 24th 2007
Serial Numbers: 99060001-99062554 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
12/07/2012 |
Responsible Entity |
|
Reason/Issue |
Partial insertion of the power cord on the RMS power input module can cause the plug to overheat and potentially create a fire hazard. Dimension Reagent Management Systems without a power cord retention clip may be affected. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Siemens is providing work around instructions for users to implement until a permanent RMS power cord retention clip can be fitted. |