Print version of this report
Type of Product Medical Device
TGA Recall Reference RC-2012-RN-00694-3
Product Name/Description CT1133S Cephalexin 100ug CL100 Antimicrobial Susceptibility Discs (Used for antimicrobial susceptibility testing), An in vitro diagnostic medical device (IVD)

Batch Number: 1129970
Recall Action Level Hospital
Recall Action Classification Class II
Recall Action Commencement Date 12/07/2012
Responsible Entity Oxoid Australia Pty Ltd
Reason/Issue The manufacturer's investigations confirmed that some cartridges within this batch may contain individual discs that are not sufficiently dosed with antibiotic. Continued use of this lot could therefore result in false indication of resistance.
Recall Action Recall
Recall Action Instructions Remove all affected lots from use and return to Oxoid Australia. This action has been closed-out on 02/02/2016.
Contact Information 08 8238 9044 - Oxoid Australia Regulatory Affairs