Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-00681-3 |
Product Name/Description |
Birdie, Birdie Compact and Flamingo Lifters (Patient Lifters)
All Model Numbers
ARTG Number: 166395 |
Recall Action Level |
Consumer |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
20/07/2012 |
Responsible Entity |
|
Reason/Issue |
Issue 1: The carabineer used to attach the spreader bar to the lifter may show some wear in the case of intensive or severe use. Issue 2: During operation by customers, the spreader bar may unintentionally unhook from the carabineer. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Issue 1: Invacare has informed that they will check and replace the carabineer if the thickness of its sections measures less than 6mm.
Issue 2: Invacare has developed retrofit kits and are intending to address this issue by reworking all the affected units. |