Type of Product |
Medical Device |
TGA Recall Reference |
RC-2012-RN-00670-3 |
Product Name/Description |
Elekta Monaco rel. 3.10.00 through 3.20.00 & Focal rel. 4.64.00 through 4.70.00 (Radiation Therapy Treatment Planning System)
Elekta Monaco rel. 3.10.00 through 3.20.00 ARTG Number: 186856
Focal rel. 4.64.00 through 4.70.00 ARTG Number: 186853 |
Recall Action Level |
Hospital |
Recall Action Classification |
Class II |
Recall Action Commencement Date |
5/07/2012 |
Responsible Entity |
|
Reason/Issue |
IF the DICOM coordinates option is turned on in the Reports pull down menu the following issues are seen on the resulting print out: Issue 1: The coordinates will be in DICOM space, thus not matching the Setup Reference dialogue. Issue 2: The shift coordinates are reported incorrectly. The Z coordinate should become DICOM negative Y and the Y coordinate should become DICOM Z. Incorrect shift coordinates are generated based on Issue 2 above. These shifts can lead to incorrect patient positioning prior to treatment position verification. |
Recall Action |
Recall for Product Correction |
Recall Action Instructions |
Customers to implement the work around instructions provided by the sponsor until the problem is resolved with the future release of a software patch. This action has been closed-out on 12/02/2016. |
Contact Information |
02 8907 1800 - Elekta Regulatory Affairs |