TGA Action ID RC-2026-RN-00968-1
Type of Product Medical Device
Product Name/Description ID-DiaClon ABO/D + Reverse Grouping for Patients. An in vitro diagnostic medical device (IVD).

Catalogue Number: 001265

Batch Number: 0962067154

Expiry date: 31/03/2027
ARTG Number 212527
Unique Device Identifier (UDI) 07611969071623
Action Type Product Alert
Action Level Hospital
Hazard Classification Class III
Reason for the Action Bio-Rad has identified unexpected positive or weak-positive reactions in the control (ctl) microtube of ID-DiaClon ABO/D + Reverse Grouping for Patients cards from Batch 0962067154. These reactions have been observed when testing RhD-positive patient samples and RhD-positive quality control materials, both manually and using IH-500 and IH-1000 instruments. Per the Instructions for Use (IFU B001265 v.01 2020-11-05), a positive reaction in the ctl microtube invalidates the test result and requires repeat testing.

Affected cards may result in invalid tests requiring repeat testing and may lead to delays in obtaining patient blood group results. When the IFU is followed, the problem is expected to result in test invalidation rather than the reporting of erroneous ABO/RhD determination.
Instructions for Customers/Users Customers are to:
1. Identify any inventory of ID-DiaClon ABO/D + Reverse Grouping for Patients, Batch 0962067154, currently in their possession.
2. Continue to follow the Instructions for Use (IFU B001265 v.01 2020-11-05). If a positive or weak-positive reaction is observed in the control (ctl) microtube, consider the test result to be invalid and repeat the test. If the control (ctl) reaction remains positive, repeat the test using another lot. If the control (ctl) reaction remains positive, undertake further investigation in accordance with the IFU before reporting results.

Results obtained with a negative control (ctl) reaction remain valid. Retrospective retesting of patient samples is not recommended.
Action Commencement Date 24/11/2026
Responsible Entity Bio-Rad Laboratories Pty Ltd
Contact Information ra_pacific@bio-rad.com - Bio-Rad