TGA Action ID RC-2026-RN-00967-1
Type of Product Medical Device
Product Name/Description FRED 2 Anti-Fog Kit

Model Number: 220-50

Lot Numbers: P5L1010, P5L1013, P5L1016, P5L1018, P5L1019, P5L1146, P5L1147, P5L1149, P6C0693, P6C0694, P6C0695, P6C0800, P6C0829, P6C0830, P6C0831, P6C0978, P6C0981, P6C1000, P6C1002, P6D1054, P6D1056, P6D1058, P6D1061, P6D1062, P6D1063, P6D1064, P6D1069, P6D1070, P6E0673, P6E0674, P6E0675, P6E0676, P6E0677

Multiple lot numbers of kits / boxes containing the Fred 2 Anti-Fog Kit
ARTG Number 178499
Unique Device Identifier (UDI) 10884521066540
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Medtronic has identified the potential for the bottle within the FRED 2 Anti-Fog Kit to leak solution due to variability in the manufacturing process. As a result of this variability, a portion of the affected product lot numbers have been found to leak solution from the bottle. Leaked solution may come in contact with the Tyvek lid. If it permeates through the Tyvek lid, the integrity of the product’s sterile barrier may be affected, which introduces the potential for biocontamination.

Internal investigations for the affected product lot numbers have found approximately 0.064% of units to exhibit evidence of solution leaking from the bottle. Leakage of solution is apparent through liquid staining, reduced solution volume within the bottle, or wet packaging material. Medtronic has not received any confirmed reports of biocontamination.
Instructions for Customers/Users Users are advised:

· Before use, inspect the FRED 2 Anti-Fog Kit box, packaging, and contents for wetness, staining, visible leakage, compromised packaging integrity, or reduced solution volume.
o Note: If the FRED 2 Anti-Fog Kit is contained within a larger kit, such as a procedure kit, other contents within are unaffected provided their packaging is intact.

· If any of these conditions are observed or the FRED 2 Anti-Fog Kit is suspected of being wet:
o Do not use the product or introduce it into the sterile field.
o Remove the product from use and quarantine it in accordance with the hospital’s applicable procedures awaiting further guidance from Medtronic Representatives.
o File a complaint with Medtronic to document the incident and facilitate potential product return.

· If no signs are observed, the FRED 2 Anti-Fog Kit may be used in accordance with the Instructions for Use. No further action is required.
Action Commencement Date 25/11/2026
Responsible Entity Medtronic Australasia Pty Ltd
Contact Information 0402 897 508 - Samantha Lord, Medtronic Australasia