TGA Action ID RC-2026-RN-00965-1
Type of Product Medical Device
Product Name/Description IOLMaster 700

Software Version: 1.90 to 1.92.35
ARTG Number 128166
Action Type Product Correction
Action Level Hospital
Hazard Classification Class III
Reason for the Action Carl Zeiss has identified two potential problems with the use of the IOLMaster 700:

The first problem relates to network configuration. The device may not retain its network settings after a restart when configured for network use for the first time. This affects new devices and devices updated from software version 1.90 to 1.92.35 that are first connected to a network after the update. Network-dependent functions, such as data export, are unavailable until the network connection is restored.

The second problem affects DICOM/PDF export. In some cases, the device cannot create PDF files when Surgeon printout biometrics, Reference image and Analysis printout are included. The device displays an unexpected error message, shuts down automatically and does not complete the export or PDF report. Direct printing is not affected.
Instructions for Customers/Users Customers are to:
- Review affected devices and contact their local ZEISS representative if the network configuration problem occurs. ZEISS can apply a temporary workaround and will install the software update when available.
- If the DICOM/PDF export problem occurs, temporarily disable Surgeon printout biometrics, Reference image and Analysis printout, then retry the export or print directly.
Action Commencement Date 20/11/2026
Responsible Entity Carl Zeiss Pty Ltd
Contact Information 1300 365 470 - Carl Zeiss