TGA Action ID RC-2026-RN-00964-1
Type of Product Medical Device
Product Name/Description SPECTRALIS H4 Instrument Table

Serial Number HEW25041069
ARTG Number 478407
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The welding of the attachment point for the castor wheels of the table may not meet quality standards. The weld may break, causing the wheel mounting sleeve to separate from the base. This defect may result in the detachment of one or more wheels, potentially affecting the stability and safe operation of the table.

The failure may occur when the device is being pulled or pushed during transportation, or if the table is subjected to additional loading (e.g. someone leaning or sitting on the table, heavy items being placed on the table). This may lead to the risk of castor detachment, potentially resulting in loss of stability of the table. As a consequence of the instability of the table, there is a potential risk of crushing of the user's or patient's body, or damage to the table and equipment on it.

This action was undertaken prior to TGA notification.
Instructions for Customers/Users Customers are to:
- Minimise additional movement or loading of the table. Ensure operators and patients remain clear of pinch points and do not push down on the tabletop.

Heidelberg Engineering has inspected the device. This inspection will enable an assessment of the condition of the product. In specific cases, the device may be returned to the manufacturer for further technical evaluation.
Action Commencement Date 24/11/2026
Responsible Entity Heidelberg Engineering Pty Ltd
Contact Information 03 9639 2125 - Heidelberg Engineering