| TGA Action ID |
RC-2026-RN-00958-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Artisan Reticulin-Nuclear Fast Red Stain Kit and Artisan Reticulin-No Counterstain Kit. An in vitro diagnostic medical device (IVD).
Artisan Reticulin-Nuclear Fast Red Stain Kit Affected Lot Numbers: 11976087, 12005481, 11965790 and 12013022 Part Number: AR17911-2 and AR17911-5
Artisan Reticulin-No Counterstain Kit Affected Lot Numbers: 12013035 Part Number: AR18211-2 |
| ARTG Number |
196077 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Specific lots of AR17911-2, AR17911-5 and AR18211-2 have been reported to exhibit weak, absent, or inconsistent staining.
The use of required run controls and review by a qualified pathologist alongside morphology, clinical history, and other diagnostic information makes this problem highly detectable.
There is the potential risk of false negative results, potential delay in testing or potential re-biopsy if tissue is exhausted. |
| Instructions for Customers/Users |
Customers are to: 1. Check inventory and confirm the number of affected kits and discard them as per local policy. 2. Indicate the number of requested replacements, if needed, on the Customer Response Form. 3. Replacement kits will then be provided by Agilent. |
| Action Commencement Date |
18/11/2026 |
| Responsible Entity |
|
| Contact Information |
enia.kakaflikas@agilent.com - Agilent Technologies Australia |