| TGA Action ID |
RC-2026-RN-00956-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
ABL90 FLEX and ABL90 FLEX PLUS Analysers running Windows XPE, Windows 7, and Windows 10. An in vitro diagnostic medical device (IVD).
Software Version: 3.5 MR11 and below |
| ARTG Number |
228382 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class III |
| Reason for the Action |
Analyser may cease to perform calibrations upon restart.
This is a follow-up on the communications distributed in June (RC-2025-RN-00534-1) and December 2025 (RC-2025-RN-01083-1).
Radiometer has completed the investigation and is implementing a permanent countermeasure via a supplement to the Instructions for Use (IFU) for Analysers running with Windows XPE, Windows 7, and Windows 10 (for Windows 10, on UIM945 computers only). |
| Instructions for Customers/Users |
Customers are to follow the instructions provided in the "Supplement to the ABL90 FLEX and ABL90 FLEX PLUS Instructions for Use for Analysers running Windows XPE, Windows 7, and Windows 10 (for Windows 10, on UIM945 computers only)" and implement the recommendations described within the supplement. |
| Action Commencement Date |
19/11/2026 |
| Responsible Entity |
|
| Contact Information |
radiometeranzqra@dhdiagnostics.com - Radiometer |