Print version of this report
TGA Action ID RC-2026-RN-00952-1
Type of Product Medical Device
Product Name/Description Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analysers. An in vitro diagnostic medical device (IVD).

Multiple lot numbers and expiry dates
ARTG Number 275701
Action Type Recall
Action Level Hospital
Hazard Classification Class I
Reason for the Action Fisher Diagnostics is recalling the product Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analysers because of complaints from customers reporting approximately 6% to 15% low calibration bias after switching ammonia standard lots and attributed the bias to the calibrators.

This problem may lead to prognostic misclassification for mortality thresholds, a delay in the initiation of time-sensitive treatment measures, and/or errors in clinical treatment decisions; this may cause development of serious patient harm.
Instructions for Customers/Users Customers should determine whether they have inventory of any affected lot of Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analyzers. If affected product is identified, discontinue use immediately and dispose of the material in accordance with applicable local, regional and national regulations.

Beckman Coulter will provide replacement products for affected products identified
Action Commencement Date 24/11/2026
Responsible Entity Beckman Coulter Australia Pty Ltd
Contact Information 1800 060 878 - Customer Administration Services