| TGA Action ID |
RC-2026-RN-00952-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analysers. An in vitro diagnostic medical device (IVD).
Multiple lot numbers and expiry dates |
| ARTG Number |
275701 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class I |
| Reason for the Action |
Fisher Diagnostics is recalling the product Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analysers because of complaints from customers reporting approximately 6% to 15% low calibration bias after switching ammonia standard lots and attributed the bias to the calibrators.
This problem may lead to prognostic misclassification for mortality thresholds, a delay in the initiation of time-sensitive treatment measures, and/or errors in clinical treatment decisions; this may cause development of serious patient harm. |
| Instructions for Customers/Users |
Customers should determine whether they have inventory of any affected lot of Infinity Ammonia Reagent for Beckman Coulter AU Chemistry Analyzers. If affected product is identified, discontinue use immediately and dispose of the material in accordance with applicable local, regional and national regulations.
Beckman Coulter will provide replacement products for affected products identified |
| Action Commencement Date |
24/11/2026 |
| Responsible Entity |
|
| Contact Information |
1800 060 878 - Customer Administration Services |