| TGA Action ID |
RC-2026-RN-00945-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Remel Haemophilus Influenzae Type a 2ml/vL. An in vitro diagnostic medical device (IVD)
Product Code: R30166001
Batch Number: 6134913 |
| ARTG Number |
235676 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class III |
| Reason for the Action |
Certain lots of Haemophilus Influenzae Type A are not performing to the Instructions for Use (IFU) criteria, not positive to ATCC9006 within 60 Seconds. Failing to agglutinate with homologous cultures within a minute.
Haemophilus influenzae agglutinating sera are intended for use in slide agglutination tests to identify serologically the type of antigen of pathogenic strains of H. influenzae (type A) for epidemiological and diagnostic purposes.
There is the potential for false negative results, however, patient management should not be affected by a delay to identify the correct serogroup. |
| Instructions for Customers/Users |
Customers are to: - Quarantine affected product to prevent further use. - Destroy the affected product as per local policy or return it to Thermo Fisher Scientific to reconcile the recall.
Review the customer letter with authorised staff / Medical Director to determine the appropriate course of action, including reviewing previously generated results for any anomalies in line with local procedures. |
| Action Commencement Date |
17/11/2026 |
| Responsible Entity |
|
| Contact Information |
1300 736 767 - Thermo Fisher Scientific |