TGA Action ID RC-2026-RN-00945-1
Type of Product Medical Device
Product Name/Description Remel Haemophilus Influenzae Type a 2ml/vL. An in vitro diagnostic medical device (IVD)

Product Code: R30166001

Batch Number: 6134913
ARTG Number 235676
Action Type Recall
Action Level Hospital
Hazard Classification Class III
Reason for the Action Certain lots of Haemophilus Influenzae Type A are not performing to the Instructions for Use (IFU) criteria, not positive to ATCC9006 within 60 Seconds. Failing to agglutinate with homologous cultures within a minute.

Haemophilus influenzae agglutinating sera are intended for use in slide agglutination tests to identify serologically the type of antigen of pathogenic strains of H. influenzae (type A) for epidemiological and diagnostic purposes.

There is the potential for false negative results, however, patient management should not be affected by a delay to identify the correct serogroup.
Instructions for Customers/Users Customers are to:
- Quarantine affected product to prevent further use.
- Destroy the affected product as per local policy or return it to Thermo Fisher Scientific to reconcile the recall.

Review the customer letter with authorised staff / Medical Director to determine the appropriate course of action, including reviewing previously generated results for any anomalies in line with local procedures.
Action Commencement Date 17/11/2026
Responsible Entity Thermo Fisher Scientific Australia Pty Ltd
Contact Information 1300 736 767 - Thermo Fisher Scientific