TGA Action ID RC-2026-RN-00944-1
Type of Product Medical Device
Product Name/Description 3Dimensions and Selenia Dimensions Mammography Systems

Model Numbers: SDM-SYS-6000-2D, SDM-SYS-6000-3D, SDM-00001-2D, SDM-00001-3D, SDM-00001-M2D, SDM-00001-M3D, SDM-05000-2A3, SDM-05000-3D2, SDM-05000-3D3, SDM-05000-3DC, SDM-SYS-9000-2D, SDM-SYS-9000-3D, 3DM-SYS-INTL2D, 3DM-SYS-INTL2D-MOB, 3DM-SYS-STD
ARTG Number 166493
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Hologic has identified the potential risk of sudden, unexpected freefall due to significant wear on the vertical lift nut of the C-Arm assembly on 3Dimensions and Selenia Dimensions systems.

There is minimal impact to device function on systems with typical wear on the lead screw. However, significant wear can cause symptoms such as seizure of vertical travel, loud noises/screeching or abnormal shaking/vibration during vertical positioning. In cases where the lead screw has significant wear, freefall of the C-Arm could occur, which could cause injury to the patient and/or user.

As of October 1, 2026, Hologic has received 5 complaints of total system failure since 2019. There have been no reports of serious injury to users or patients.
Instructions for Customers/Users Customers may continue to use their systems while awaiting an inspection by Hologic. If at any time during system operation customers observe symptoms such as seizure of vertical travel, unexpected movement/shaking or hear a loud screeching during vertical travel, discontinue use and contact Hologic’s Technical Support immediately.

Hologic will perform a recurring maintenance inspection to properly identify lift nut wear. For systems under warranty or service contract, the inspection will be performed during the next schedule maintenance visit. For systems not under warrant or service contract, Hologic will contact them to schedule an inspection.
Action Commencement Date 17/11/2026
Responsible Entity Hologic (Australia & New Zealand) Pty Ltd
Contact Information 1800 264 073 - Hologic Technical Support