| TGA Action ID |
RC-2026-RN-00943-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
LP(a) Calibrator Series. An in vitro diagnostic medical device (IVD).
Catalogue Number: LP3404
Batch Number: 899841
Lot Numbers: 1955LP, 1956LP, 1957LP, 1958LP, 1959LP
Expiry Date: 28th March 2029 |
| ARTG Number |
199617 |
| Unique Device Identifier (UDI) |
05055273204438 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
A limited number of vials are cloudy in appearance and the level of LP(a) in these vials is impacted. If a calibration is performed using one of the impacted calibrators the calibration curve will be impacted.
In some cases, calibration may fail, but calibrations may also pass and patient recovery may be impacted. Due to the variation between vials and subsequent impact on calibration and patient result, customers should discontinue use of this calibrator set with immediate effect. |
| Instructions for Customers/Users |
Customers should: · Discontinue use of and discard any of the above immediately. · Review your reagent inventory of these products and assess your laboratories needs for reimbursement for discarded inventory. · Review results generated with the affected batches in line with the clinical profile of the patient. · Discuss the contents of this notice with your Medical Director to determine the appropriate course of action, including for any previously generated results. If applicable follow your laboratories protocol regarding the need for reviewing previously reported patient results. |
| Action Commencement Date |
16/11/2026 |
| Responsible Entity |
|
| Contact Information |
02 9615 4640 - Randox Australia |