| TGA Action ID |
RC-2026-RN-00940-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
BD Connecta Stopcock with Extension Tube - white 16 cm
Product Code: 394995
Lot Numbers: 5004293, 5009468, 5031115, 5042456, 5073966, 5087792, 5120823, 5120824, 5126986, 5132786, 5132787, 5244329, 5275634, 5275637, 5275640, 5331911, 6003373 |
| ARTG Number |
128272 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Specific lot numbers of the 394995 BD Connecta Stopcock with Extension Tube may have been manufactured with an inadequate amount of solvent which could result in separation of the tubing from the male luer component.
All other active Connecta Stopcock catalogue numbers are not affected by this problem.
Separation of the Connecta Stopcock with Extension Tube may occur during use, resulting in a compromised fluid pathway. This may result in leakage with the potential consequences including under-infusion or disruption of prescribed therapy, exposure to potentially hazardous fluids, infection or air embolism.
The separation may be apparent upon initiation of infusion, or during standard patient care at which point the healthcare professional should promptly discontinue use and replace the affected stopcock extension set.
To date, in Australia, BD has received zero (0) reports of this problem, and no adverse events have occurred. |
| Instructions for Customers/Users |
Customers are advised:
1. If the product is currently in use, replace immediately with unaffected product. - If unable to determine if the product is within scope, it is recommended to replace with unaffected product.
2. If affected product has already been used and is no longer in situ, no patient follow up activities are required.
3. Please discontinue use of the affected lot numbers and immediately dispose of all affected product in accordance with each facility’s process. |
| Action Commencement Date |
16/11/2026 |
| Responsible Entity |
|
| Contact Information |
1800 656 100 - Customer Service team |