TGA Action ID RC-2026-RN-00932-1
Type of Product Medical Device
Product Name/Description Polaris Comfort, Pro and Pro+ Surgical Light

Product Code: GG16100
ARTG Number 476167
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Under certain operating conditions, a signal interruption may occur between the LED module and the control board.

In such a case, the surgical light automatically switches to a safe operating mode. The lighting function remains active, but the controls cannot be operated while in this state.

There is the potential for procedural complications if the operating light cannot be controlled during surgery. The probability of an injury is considered improbable, because the device enters its defined fail-safe state and continues to provide =40 klx illumination, and ORs are equipped with redundant lighting systems.
Instructions for Customers/Users Draeger will carry out a replacement of your Polaris surgical light heads. The replacement is carried out by qualified specialist personnel.

Until then, customer may continue to operate the devices. lf the error occurs, it may be necessary to switch to the redundant OR light head.
Action Commencement Date 16/11/2026
Responsible Entity Draeger Australia Pty Ltd
Contact Information 1800 372 437 - Draeger Australia Customer Service