| TGA Action ID |
RC-2026-RN-00932-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Polaris Comfort, Pro and Pro+ Surgical Light
Product Code: GG16100 |
| ARTG Number |
476167 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Under certain operating conditions, a signal interruption may occur between the LED module and the control board.
In such a case, the surgical light automatically switches to a safe operating mode. The lighting function remains active, but the controls cannot be operated while in this state.
There is the potential for procedural complications if the operating light cannot be controlled during surgery. The probability of an injury is considered improbable, because the device enters its defined fail-safe state and continues to provide =40 klx illumination, and ORs are equipped with redundant lighting systems. |
| Instructions for Customers/Users |
Draeger will carry out a replacement of your Polaris surgical light heads. The replacement is carried out by qualified specialist personnel.
Until then, customer may continue to operate the devices. lf the error occurs, it may be necessary to switch to the redundant OR light head. |
| Action Commencement Date |
16/11/2026 |
| Responsible Entity |
|
| Contact Information |
1800 372 437 - Draeger Australia Customer Service |