TGA Action ID RC-2026-RN-00927-1
Type of Product Medical Device
Product Name/Description Philips Incisive CT system

Software Versions: V4.5 and/or V5.0

Product Code (REF): 728143, 728144
ARTG Number 352131
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Philips has identified two potential problems with the Incisive CT system with software V4.5 and/or V5.0

Problem 1:
This may occur on Incisive CT with software V5.0 and configured with the Precise Position function:
When using Precise Position, the system displays the patient orientation and surview range on the console and gantry panel. In rare cases, after the operator clicks “Start Exam,” the patient orientation information may be cleared in the background process, and the patient orientation icon may disappear from the screen. If the operator does not notice this and continues with the surview scan, the surview image may be acquired from the wrong orientation or position.

Problem 2:
This may occur on Incisive CT with software V4.5&V5.0 during Bolus Tracking in manual ROI workflow: During Bolus Tracking, the system monitors the contrast level within the selected Region of Interest (ROI) and is designed to automatically trigger the clinical scan when the predefined contrast threshold is reached. Under specific mouse-button operations in the tracker interface—pressing the left mouse button and the middle or right mouse button either simultaneously or in rapid succession, an invalid, invisible automatic ROI may be created. If the operator then switches to the manual ROI workflow and manually adds ROIs, the CT value in the timing diagram may not be updated. As a result, the system may not automatically trigger the clinical scan when the contrast threshold is reached.
Instructions for Customers/Users Philips will contact customers to schedule a time for a field engineer to correct the problems. In the interim, customers may continue to use the system(s), by following these recommendations:

Problem 1 - After clicking “Start Exam” but before proceeding with surview acquisition using the Precise Position workflow, verify that the patient orientation displayed on the console and gantry panel is correct. If the patient orientation indication icon is absent, or if the displayed orientation does not match the intended patient orientation, do not proceed with the surview scan. Manually resolve the patient orientation information before continuing the examination.

Problem 2 - During Bolus Tracking, Philips recommends using the default automatic ROI workflow. If use of the manual ROI workflow is necessary, avoid pressing the left mouse button and the middle or right mouse button simultaneously or in rapid succession in the tracker interface.
Action Commencement Date 12/11/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team