TGA Action ID RC-2026-RN-00926-1
Type of Product Medical Device
Product Name/Description Xpert MRSA/SA SSTI. An in vitro diagnostic medical device (IVD).

Batch Number: 1001489295

Part Number: GXMRSA/SA-SSTI-CE

Lot Number: 17701

Expiry Date: 04/07/2027
ARTG Number 242177
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Internal testing identified the occurrence of false-negative results in positive samples tested using GXMRSA/SA-SSTI-CE batch 1001489295 (Lot No 17701).

False-negative results, such as MRSA NEGATIVE/ SA POSITIVE or MRSA NEGATIVE/ SA NEGATIVE, may contribute to inappropriate or delayed treatment. In some cases, this could result in progression of infection, potentially including bloodstream infection and sepsis.
Instructions for Customers/Users Customers should:
1. Stop using any GXMRSA/SA-SSTI batch 1001489295 (lot 17701) kits at their facility and dispose of the remaining affected product
2. Complete the Customer Response Form to acknowledge receipt and request replacement product.
3. As with all questionable negative or questionable positive results, please report to Cepheid Technical Support for investigation and the TGA via a Medical Device Incident Report.
Action Commencement Date 10/11/2026
Responsible Entity Cepheid Holdings Pty Ltd
Contact Information 1800 130 821 - Cepheid Holdings