TGA Action ID RC-2026-RN-00915-1
Type of Product Medical Device
Product Name/Description Mini-Blades and Lamellar Blades

Beaver Mini-Blade (rounded tip sharp on one side)
SKU: 364820

Beaver Mini-Blade Contains 100 Blades BEAVER6700
SKU: 364822

Beaver Mini-Blade Contains 100 Blades BEAVER6900
SKU: 364823

Beaver Mini-Blade (rounded tip sharp on one side)
SKU: BEAVER6400

Beaver Lamellar Blade, sharp all around
SKU: BEAVER6700

Beaver Mini-Blade
SKU: BEAVER6900

Multiple lot numbers
ARTG Number 181674
Unique Device Identifier (UDI) 00886158004455, 00886158004479, 00886158004486, 00886158004394, 00886158004417, and 00886158004424
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Beaver-Visitec International (BVI) identified a seal integrity problem affecting certain units. The problem may result in an ‘open seal’ condition, in which sterility cannot be guaranteed.

An open seal compromises the integrity of the sterile barrier and results in loss of product sterility. Use of a product with an open seal may therefore expose the patient to a non-sterile product and may result in an increased risk of infection.

To date, BVI has received one complaint indicating that multiple products had been found with an open seal and has not received any reports of adverse health outcomes related to this problem.
Instructions for Customers/Users Customers are to:
- Review their existing stock of the affected lots.
- Place under quarantine identified SKUs and lots, to not use or further distribute product subject to this recall notification.
- Indicate in the customer response form how many units have been quarantined.

The sponsor will contact customers with further instructions regarding the product return process.
Action Commencement Date 11/11/2026
Responsible Entity Beaver-Visitec International Australia Pty Ltd
Contact Information carmen@mcfarlanemedical.com.au - McFarlane Medical