| TGA Action ID |
RC-2026-RN-00912-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Liquichek Unassayed Chemistry Control, Liquid Unassayed Multiqual, Liquid Assayed Multiqual & Liquid Assayed Multiqual Premium. An in vitro diagnostic medical device (IVD).
Multiple catalogue numbers, lot numbers and expiry dates |
| ARTG Number |
198600 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Bio-Rad has identified a stability problem affecting the quality control material lots listed above. Investigation found that Triglycerides, Phosphorus, and Magnesium values increase over time after the vial is thawed and opened (open-vial stability). Magnesium values also decrease over the product’s shelf life.
Use of affected quality control lots may result in QC shifts for the impacted analytes, which could require laboratory investigation, repeat testing, or use of alternate control material with subsequent temporary operational delays related to QC investigation. |
| Instructions for Customers/Users |
Bio-Rad is requesting that customers take the following actions: • Review your inventory and identify any affected product listed by product name, catalog number, lot number, expiration date, and UDI. • Follow your laboratory’s established quality control procedures for evaluating QC performance, including investigation of out-of-control results or observed shifts for Triglycerides, Phosphorus, or Magnesium. • Please obtain a revised copy of the product insert, which now includes the revised Storage and Stability and Limitations section through the Internet at http://myeinserts.qcnet.com • Discard all previous versions of the product inserts for these lots of product that you may have at your location and substitute with the revised package inserts. • Use alternate QC material or replacement product, as applicable, if your laboratory determines the affected lot no longer meets your internal acceptance criteria. |
| Action Commencement Date |
6/11/2026 |
| Responsible Entity |
|
| Contact Information |
(02) 9914 2800 |