TGA Action ID RC-2026-RN-00910-1
Type of Product Medical Device
Product Name/Description Vercise Deep Brain Stimulation Systems

Vercise Deep Brain Stimulation Systems Clinician Programmer
Material Number: M365DB71610 & M365DB71640

Vercise Neural Navigator 5
Material Number: M365DB7105N501A0 & M365DB7105N5A0
ARTG Number 308574, 462219
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The non-rechargeable IPG battery longevity estimate (provided as the Energy Use Index [EUI]) and rechargeable IPG battery recharge estimate are supplemental informational tools intended to support IPG replacement and recharge planning, respectively. These estimates are not intended to provide real-time ongoing monitoring of battery status.

Boston Scientific is updating the Programming Manual Instructions for Use to provide additional guidance in the use of the Clinician Programmer feature for Implantable Pulse Generator (IPG) battery recharge interval and non-rechargeable IPG battery longevity estimates.

If the actual impedance of the active contacts are higher or lower than 1,000 ohms(?), the battery longevity or recharge estimate will decrease or increase accordingly relative to the actual impedance values. An inaccurate estimate does not interrupt therapy or directly cause patient harm. However, if this results in delayed battery replacement or inadequate recharging, subsequent battery depletion could result in interruption of stimulation.
Instructions for Customers/Users Customers are to:
- Review the updates to the Programming Manual Instructions for Use provided with the customer letter.
Action Commencement Date 9/11/2026
Responsible Entity Boston Scientific Pty Ltd