| Reason for the Action |
When two non-defibrillation-proof transducers are used, the system fails to maintain defibrillation protection because the adapter lacks connector pin spacing, and therefore the 12-pin connector cannot withstand the required 5 kV for defibrillation protection. In clinical situations where defibrillation is needed the use of non-defibrillation-proof transducers can create hazardous situations where defibrillation energy may be partially shunted through the IBP system rather than fully through the patient.
For the problem to occur, the Dual IBP Adapter must be used with transducers not listed as compatible in the Instructions for Use (IFU). As a result, use of the adapter with catheter-tip transducers particularly increases risk, as these transducers are typically not defibrillation-proof due to limitations inherent to their design. The Dual IBP Adapter design does not prevent use with non-defibrillation-proof transducers.
When this problem occurs, patients may be exposed to the operational hazard of “Loss of Function” caused by the delivery of insufficient defibrillation energy. In addition, this problem may cause the monitor to not recover fast enough, or at all, after defibrillation, which can expose patients to incorrect or delayed treatment. If the adapter is touched by the user during defibrillation, the user may be exposed to an electrical hazard which can cause electrical shock and contribute to death. |