TGA Action ID RC-2026-RN-00908-1
Type of Product Medical Device
Product Name/Description Dual IBP Adapter

Model Number: 989803199741
ARTG Number 292264
Unique Device Identifier (UDI) 00884838085299
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action When two non-defibrillation-proof transducers are used, the system fails to maintain defibrillation protection because the adapter lacks connector pin spacing, and therefore the 12-pin connector cannot withstand the required 5 kV for defibrillation protection. In clinical situations where defibrillation is needed the use of non-defibrillation-proof transducers can create hazardous situations where defibrillation energy may be partially shunted through the IBP system rather than fully through the patient.

For the problem to occur, the Dual IBP Adapter must be used with transducers not listed as compatible in the Instructions for Use (IFU). As a result, use of the adapter with catheter-tip transducers particularly increases risk, as these transducers are typically not defibrillation-proof due to limitations inherent to their design. The Dual IBP Adapter design does not prevent use with non-defibrillation-proof transducers.

When this problem occurs, patients may be exposed to the operational hazard of “Loss of Function” caused by the delivery of insufficient defibrillation energy. In addition, this problem may cause the monitor to not recover fast enough, or at all, after defibrillation, which can expose patients to incorrect or delayed treatment. If the adapter is touched by the user during defibrillation, the user may be exposed to an electrical hazard which can cause electrical shock and contribute to death.
Instructions for Customers/Users Users should ensure that compatible disposable pressure transducers and accessories (per IFU) are used in conjunction with the Dual IBP Adapter.

A Philips representative will send a follow-up communication to provide instructions on how to obtain updated Instructions for Use (IFU) and Addendum once available.
Action Commencement Date 16/11/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Service Delivery Team