| Product Name/Description |
RIDA Xplore System and HighPlex System. An in vitro diagnostic medical device (IVD).
Software: RIDA Setup 3.01.05
Product Codes: 91501 and 96521 |
| ARTG Number |
177847 |
| Unique Device Identifier (UDI) |
09343044002298 , 09343044005848 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
A software defect was identified in RIDA Setup v3.01.05, affecting the RIDA Xplore System (96521) and HighPlex System (91501). The defect occurs only when all the following conditions are met: Samples are entered manually; the POSCON option is selected; and the positive control is placed in any position other than the last position in the run.
Under these conditions, the software repositions a sample without updating the corresponding sample identifier, resulting in a mismatch between the physical sample position and the assigned sample label. This creates the potential for incorrect sample identification and assignment of results to the wrong specimen. |
| Instructions for Customers/Users |
Customers are to be aware: - Not to tick the POSCON option and use positive control in last position. - As an interim risk control, the IFU will be updated to advise users not to operate the device in the identified manner until a software correction has been implemented. |