TGA Action ID RC-2026-RN-00898-1
Type of Product Medical Device
Product Name/Description RIDA Xplore System and HighPlex System. An in vitro diagnostic medical device (IVD).

Software: RIDA Setup 3.01.05

Product Codes: 91501 and 96521
ARTG Number 177847
Unique Device Identifier (UDI) 09343044002298 , 09343044005848
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action A software defect was identified in RIDA Setup v3.01.05, affecting the RIDA Xplore System (96521) and HighPlex System (91501). The defect occurs only when all the following conditions are met: Samples are entered manually; the POSCON option is selected; and the positive control is placed in any position other than the last position in the run.

Under these conditions, the software repositions a sample without updating the corresponding sample identifier, resulting in a mismatch between the physical sample position and the assigned sample label. This creates the potential for incorrect sample identification and assignment of results to the wrong specimen.
Instructions for Customers/Users Customers are to be aware:
- Not to tick the POSCON option and use positive control in last position.
- As an interim risk control, the IFU will be updated to advise users not to operate the device in the identified manner until a software correction has been implemented.
Action Commencement Date 4/11/2026
Responsible Entity AusDiagnostics Pty Ltd
Contact Information 02 9698 8030 - AusDiagnostics Pty Ltd