TGA Action ID RC-2026-RN-00895-1
Type of Product Medical Device
Product Name/Description InteleShare (cloud-hosted; no client-installed component)

All releases from 3.25.2.0 onward – 3.25.2, 3.26.1, 3.26.2, and 3.26.3 lines.
ARTG Number 461485
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The sponsor identified two software problems affecting certain versions of InteleShare.

The first problem may result in an incomplete study being duplicated when images are still being received, as subsequent copy requests may fail to transfer the remaining images.

The second problem may result in DICOM images being removed from primary storage without being preserved in the archive during study archiving or deletion. In either case, a study may appear complete or successfully transferred to PACS despite images being missing.
Instructions for Customers/Users Customers are to:
- Verify study completeness, including source and destination image counts where duplicate images are used, and do not rely on studies where completeness cannot be confirmed.
- Review patient records from 7 October 2025 where duplicate images have been used, if clinically appropriate, and follow the instructions provided in the customer letter for managing suspected incomplete studies.
- Allow Intelerad to update all affected products.
Action Commencement Date 4/11/2026
Responsible Entity Emergo Asia Pacific Pty Ltd T/a Emergo Australia
Contact Information AUNZvigilance@ul.com - Emergo Australia