TGA Action ID RC-2026-RN-00882-1
Type of Product Medical Device
Product Name/Description GR3 Vital Signs Monitor

Model: G3R

Product Code: 4604
ARTG Number 518246
Unique Device Identifier (UDI) 06938368100244
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The GR3 Vital Signs Monitor has been supplied with mains power cords that do not comply with AS/NZS 3112, as the plugs do not have insulated active and neutral pins.

This presents a potential electric shock hazard if a person comes into contact with a live pin while inserting or removing the plug from a power outlet. Electric shock may result in injury.
Instructions for Customers/Users Customers are to:
- Stop using the affected power cords and disconnect them from the mains supply when it is safe to do so. Ensure the affected power cords are not used until they have been replaced.

Cubro will arrange for a representative to replace the non-compliant power cords with compliant, manufacturer-approved replacement cords.

Cubro is continuing to investigate the full extent of affected shipments and will provide an update if additional affected units are identified.

This action has been closed out on 12/11/2026
Action Commencement Date 2/11/2026
Responsible Entity Cubro Pty Ltd