| TGA Action ID |
RC-2026-RN-00882-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
GR3 Vital Signs Monitor
Model: G3R
Product Code: 4604 |
| ARTG Number |
518246 |
| Unique Device Identifier (UDI) |
06938368100244 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
The GR3 Vital Signs Monitor has been supplied with mains power cords that do not comply with AS/NZS 3112, as the plugs do not have insulated active and neutral pins.
This presents a potential electric shock hazard if a person comes into contact with a live pin while inserting or removing the plug from a power outlet. Electric shock may result in injury. |
| Instructions for Customers/Users |
Customers are to: - Stop using the affected power cords and disconnect them from the mains supply when it is safe to do so. Ensure the affected power cords are not used until they have been replaced.
Cubro will arrange for a representative to replace the non-compliant power cords with compliant, manufacturer-approved replacement cords.
Cubro is continuing to investigate the full extent of affected shipments and will provide an update if additional affected units are identified.
This action has been closed out on 12/11/2026 |
| Action Commencement Date |
2/11/2026 |
| Responsible Entity |
|
| Contact Information |
1800 656 527 - Cubro Pty Ltd |