Print version of this report
TGA Action ID RC-2026-RN-00879-1
Type of Product Medical Device
Product Name/Description Rotaflow Console

RotaFlow English/UK UK-Plug ICU
Item Number: 701051697

MCP00706041#RotaFlow English/UK UK-Plug
Item Number: 701043291

RotaFlow English/Australia-Plug
Item Number: 701053966

MCP00706037#RotaFlow English EU-Plug
Item Number: 701043290

MCP00706051#RotaFlow German EU-Plug
Item Number: 701043294
ARTG Number 179104
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Investigation found that, after the ROTAFLOW Console operates in battery mode and reconnects to external power, a software problem may cause battery charging to stop prematurely. The charging indicator turns off even though the battery is not fully charged, which may leave the battery with less available operating time than expected.

No patient injuries or deaths associated with this problem have been reported.
Instructions for Customers/Users Customers are to:
- Check inventory to identify affected devices.
- During patient support: After reconnecting the device to mains power following battery operation, check the charging indicator after 3 minutes. If the indicator is off, reconnect the mains power and repeat the check after 3 minutes.
- When the device is not being used for patient support: Switch the ROTAFLOW Console off, disconnect it from mains power, wait 30 seconds, then reconnect it to mains power and verify that the charging indicator is illuminated.
Action Commencement Date 16/11/2026
Responsible Entity Getinge Australia Pty Ltd
Contact Information 1800 438 464 - Customer Service