| TGA Action ID |
RC-2026-RN-00879-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Rotaflow Console
RotaFlow English/UK UK-Plug ICU Item Number: 701051697
MCP00706041#RotaFlow English/UK UK-Plug Item Number: 701043291
RotaFlow English/Australia-Plug Item Number: 701053966
MCP00706037#RotaFlow English EU-Plug Item Number: 701043290
MCP00706051#RotaFlow German EU-Plug Item Number: 701043294 |
| ARTG Number |
179104 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Investigation found that, after the ROTAFLOW Console operates in battery mode and reconnects to external power, a software problem may cause battery charging to stop prematurely. The charging indicator turns off even though the battery is not fully charged, which may leave the battery with less available operating time than expected.
No patient injuries or deaths associated with this problem have been reported. |
| Instructions for Customers/Users |
Customers are to: - Check inventory to identify affected devices. - During patient support: After reconnecting the device to mains power following battery operation, check the charging indicator after 3 minutes. If the indicator is off, reconnect the mains power and repeat the check after 3 minutes. - When the device is not being used for patient support: Switch the ROTAFLOW Console off, disconnect it from mains power, wait 30 seconds, then reconnect it to mains power and verify that the charging indicator is illuminated. |
| Action Commencement Date |
16/11/2026 |
| Responsible Entity |
|
| Contact Information |
1800 438 464 - Customer Service |