| TGA Action ID |
RC-2026-RN-00876-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
4132 Hydragel 15 ISO-PAL. An in vitro diagnostic medical device (IVD).
Lot Numbers: 03066/51, 24046/51
Expiry Dates: 31/01/2028, 31/01/2028 |
| ARTG Number |
197918 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class III |
| Reason for the Action |
A gradual decrease in the performance of Lectin lot 26115/01 has been identified during the stability monitoring performed on this lot.
This decrease in performance was first observed after one month under accelerated stability conditions (stress testing to simulate and predict long-term degradation) and was subsequently confirmed during evaluations conducted at later time points.
To date, under normal storage and use conditions, the reagent remains compliant. However, the stability results suggest that a loss of performance may occur before the currently claimed expiry date. |
| Instructions for Customers/Users |
Customers are to: - Inspect their stock and quarantine the affected lots to prevent further use. - Destroy the affected stock and contact the sponsor about replacements. Alternative lots are currently available for order. |
| Action Commencement Date |
29/10/2026 |
| Responsible Entity |
|
| Contact Information |
1800 498 588 - Sebia Australia |