| TGA Action ID |
RC-2026-RN-00872-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Philips Azurion Systems
Azurion 7 B12 System Code: 722067
Azurion 7 B20 System Codes: 722068, 722226
Azurion 7 M12 System Codes: 722078, 722223
Azurion 7 M20 System Codes: 722079, 722224
Azurion 5 M20 System Code: 722228 |
| ARTG Number |
225815, 370112 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Philips has identified that, during certain repair or service actions, Solid-State Drives (SSDs) that are not specified in the Medical Device File may have been installed. These SSDs may have different performance characteristics, which may include reduced reliability over time.
This may result in loss of imaging functionality or motorised movement, a blank or black display, and/or the inability to store newly acquired images or access previously acquired images.
This problem may potentially result in or contribute to a delay in diagnosis, delay in therapy/treatment, and/or procedural complications.
To date Philips has not received any reports of harm associated with this problem. |
| Instructions for Customers/Users |
Affected systems may continue to be used in accordance with their Instructions for Use (IFU) and the instructions in the customer letter. If motorised movements become unavailable, the stand can be manually repositioned using the handgrips and brake controls located on both sides of the stand.
Customers are to: o Contact their local Philips representative if they experience the problem.
Philips will inspect all systems in scope of this Urgent Product Correction and replace SSDs that are not specified in the Medical Device File. Philips expects to start the implementation of this correction in December 2026.
A local Philips representative will contact customers to schedule a visit for a Field Service Engineer to perform these activities. |
| Action Commencement Date |
30/10/2026 |
| Responsible Entity |
|
| Contact Information |
1800 251 400 - Philips Service Delivery Team |