TGA Action ID RC-2026-RN-00868-1
Type of Product Medical Device
Product Name/Description Brilliance MRSA 2 agar plates. An in vitro diagnostic medical device (IVD).

Item Number: PP2475

Batch Number: 4567009

Expiry Date: 20/10/2026
ARTG Number 214128
Action Type Recall
Action Level Hospital
Hazard Classification Class III
Reason for the Action Investigation found that a raw material picking error resulted in an incorrect dehydrated media base being used to manufacture a batch of Brilliance MRSA 2 agar plates.

The affected plates may fail to produce the expected denim blue colonies for MRSA and may allow breakthrough growth of Enterococcus that resembles MRSA, potentially resulting in false negative or false positive results.
Instructions for Customers/Users Customers are to:
- Destroy any affected product remaining in inventory or return it to Thermo Fisher Scientific.
- If the affected product has been used, review the results for any anomalies.
Action Commencement Date 27/10/2026
Responsible Entity Thermo Fisher Scientific Australia Pty Ltd