| TGA Action ID |
RC-2026-RN-00867-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
MAGNETOM MR Scanners and BIOGRAPH One PET/MR Scanners
Multiple models and material numbers |
| ARTG Number |
274035, 387372 |
| Action Type |
Product Correction / Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
A software problem has been identified in syngo.via MR Cardiac Analysis MR Flow Quantification affecting the evaluation of retrospectively gated 2D cardiac flow data. The algorithm does not correctly represent the final phase of the cardiac cycle. The evaluation stops at the last trigger time point (TT point), which marks the beginning of the final cardiac phase, instead of accounting for the complete duration of that phase until the end of the cardiac cycle. The final repetition-time interval (TR) is not included in the calculation (“TR missing”). This results in systematic truncation of the cardiac cycle and incomplete loop closure.
This problem can lead to an underestimation (up to 14–16%) of quantitative flow values which could contribute to an incorrect evaluation of a clinical condition and may result in a shift in diagnostic classification. |
| Instructions for Customers/Users |
Siemens Healthineers is preparing software updates to correct this problem. The applicable update will depend on the scanner model and installed software version. Not all software updates will be available at the same time. The first updates are currently planned to become available in first quarter of 2027, followed by additional updates for the remaining affected configurations.
Until a software correction is available, users should be aware that Beat Sensor-triggered examinations may result in a systematic underestimation of calculated flow values.
Flow quantification values are only one component of the overall clinical assessment. The possible underestimation needs to be considered by the physician to avoid incorrect clinical interpretation before taking final therapeutic decisions.
The manufacturer recommends using syngo.via MR Cardiac Analysis MR Flow Quantification only in combination with ECG triggering. |
| Action Commencement Date |
30/10/2026 |
| Responsible Entity |
|
| Contact Information |
1800 310 300 - Siemens Customer Care Centre |