| TGA Action ID |
RC-2026-RN-00866-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Mouse Monoclonal Primary Antibody (0.5ml conc). An in vitro diagnostic medical device (IVD).
Product Code: 326M-15 PSAP (PASE/4LJ)
Lot Number: 0000590132 |
| ARTG Number |
239099 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
An investigation identified that the lot 0000590132 of PSAP (PASE/4LJ) Mouse Monoclonal Primary Antibody may exhibit reduced staining performance, resulting in weak, absent, or no staining on positive control tissue. Customer complaints and stability testing confirmed out-of-specification performance in the affected lot.
Affected product may generate weak or negative staining results when used in immunohistochemistry (IHC) testing. This may result in: · Invalid test results. · False negative staining results. · Potential misclassification of prostate cancer origin. · Need for repeat testing and diagnostic delays. |
| Instructions for Customers/Users |
Customers are to: - Immediately inspect inventory and quarantine all affected stock to prevent further use. - Return the customer response form to replace any affected stock. |
| Action Commencement Date |
29/10/2026 |
| Responsible Entity |
|
| Contact Information |
1800 287 222 - Abacus dx Pty Ltd |