| TGA Action ID |
RC-2026-RN-00864-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
ARTIS pheno, ARTIS icono floor, ARTIS icono biplane, ARTIS icono ceiling
Software versions: VE40B & VE50B |
| ARTG Number |
273952 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Under certain conditions, the system may display an internal generator communication error after startup or restart. If this occurs, X-ray functionality may not be available, and the system may not be ready for clinical use. No unintended radiation emission has been identified in connection with this problem. This may lead to a delay, interruption, or inability to perform an intended procedure requiring X-ray imaging. |
| Instructions for Customers/Users |
The correction will be available and completed by the end of Q1 2027. Siemens will contact customers for an appointment to perform the corrective action.
Until the corrective action has been completed, users should take the possibility of system unavailability after startup or restart into account when planning clinical use of an affected system. For critical, urgent, or time-sensitive procedures, healthcare professionals should decide in advance whether the affected system should be planned for such procedures and should consider established contingency procedures, including the use of an alternative suitable system or treatment pathway, if available.
If the system is not available and an urgent or time-critical procedure is required, users should follow their established clinical contingency procedures and use an alternative suitable system or treatment pathway, if available. |
| Action Commencement Date |
29/10/2026 |
| Responsible Entity |
|
| Contact Information |
1800 310 300 - Siemens Customer Care Centre |