Type of Product Medical Device
Product Name/Description corpuls3 / corpuls3 Touch

corpuls3
Product Code: CP-04100

corpuls3 Touch
Product Code: CP-04101

Multiple affected serial numbers
ARTG Number 192745
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Printing a resting ECG may temporarily interrupt the Bluetooth connection and cause a brief display interruption. This is related to a hardware-specific problem and is not associated with other ongoing device problems such as unexpected software shutdown/reboot.

Devices manufactured between Jan 2024 and Mar 2026 are affected. The problem requires specific environmental conditions, such as low humidity or very dry printer paper.

The root cause has been identified as charge separation generated by the movement of the printer paper, which may lead to electrostatic charging of the metallic paper guide. This charge may discharge via a nearby test connection point. The resulting electrical interference can temporarily interrupt the Bluetooth connection to the monitor and consequently the display.

A temporary interruption of the display may result in users temporarily losing access to displayed patient information and being unable to operate the system for 5–25 secs. The display and system functionality are restored automatically without loss of data after re-establishment of communication.

This may lead to an interruption or delay in life-saving treatment and, in the worst case, may result in serious adverse health consequences for patients.

Between 1/1/24 and 24/09/26, there have been 67 complaints worldwide, with zero complaints received in Australia. To date, no reports of patient harm related to this failure mode have been received.
Instructions for Customers/Users Be aware that this problem can occur and use an alternative device or have a backup device available if possible.

Corpuls Pacific will replace the P509 printed circuit board (PCB) over the next 8-10 months.

All personnel, users, operators and owners of the affected device are to read and be informed about the possibility of a temporary interruption of the Bluetooth connection to the monitor, which may result in a temporary loss of the display.

The interruption is limited in duration, and the display will be restored automatically. The product can then continue to be used in accordance with its intended purpose

a. If the problem described occurs, contact Corpuls Pacific, who will replace the P509 PCB as a matter of urgency.

b. If the problem does not occur on your device, the device will be updated over the next 8-10 months.
Action Commencement Date 17/11/2026
Responsible Entity Corpuls Pacific Pty Ltd
Contact Information 1300 178 787 - Corpuls Pacific Customer Service