TGA Action ID RC-2026-RN-00859-1
Type of Product Medical Device
Product Name/Description Rabbit Monoclonal Antibody (0.1 mL Concentrate). An in vitro diagnostic medical device (IVD).

Product Code: 354R-14 TFE3 (MRQ-37)

Lot Number: 0000559553
ARTG Number 239099
Unique Device Identifier (UDI) N/A
Action Type Recall
Action Level Hospital
Hazard Classification Class III
Reason for the Action Routine stability testing identified that the 0.1 mL concentrate size of TFE3 (MRQ-37) Rabbit Monoclonal Antibody may produce weak or no staining as the product ages, prior to expiry.

Weak staining may result in an inconclusive result requiring retesting, while no staining may result in a false-negative or incorrect result.
Instructions for Customers/Users Customers are to:
- Inspect stock and identify any affected lots.
- Destroy all affected stock on site as per local policy.
- Replacement product or a credit will be provided for affected stock.
Action Commencement Date 22/10/2026
Responsible Entity Abacus dx Pty Ltd
Contact Information 1800 222 287 - Abacus dx