TGA Action ID RC-2026-RN-00817-1
Type of Product Medical Device
Product Name/Description Apple-Hunt Secondary Cannula Pyramidal Tip 5mm

Lot Number: 550034148

Catalogue Number: 900-800
ARTG Number 364354
Unique Device Identifier (UDI) 00888937003833
Action Type Recall
Action Level Hospital
Hazard Classification Class III
Reason for the Action An investigation identified that devices from the affected lot may contain a cannula component that exceeds the intended length specification. When the device is assembled in accordance with the Instructions for Use, the pyramidal trocar tip may not extend beyond the distal end of the cannula as designed.

The potential impact of this problem is limited to user inconvenience and/or a minor delay while an alternative device is obtained.

This condition is readily detectable during the required assembly and pre-use inspection process, as the trocar tip remains enclosed within the cannula following assembly.

No reports of patient injury associated with this problem have been received.
Instructions for Customers/Users Customers are to immediately inspect inventory for the affected product and quarantine all affected stock to prevent further use.

Gytech will provide replacement stock upon receipt of a completed Customer Response form.
Action Commencement Date 19/10/2026
Responsible Entity Gytech Pty Ltd
Contact Information 03 9822 5911 - Gytech