Print version of this report
TGA Action ID RC-2026-RN-00815-1
Type of Product Medical Device
Product Name/Description SIGNA Artist Evo and SIGNA Artist electronics cabinet

SIGNA Artist
GTIN: 00195278117021

SIGNA Artist Evo
GTIN: 00195278481382
ARTG Number 231238
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action GE HealthCare has become aware of an exposed sharp edge inside the SIGNA Artist Evo and SIGNA Artist electronics cabinet that could damage power cables, which may potentially cause an electrical fault. If the electrical fault is present and the cabinet door is opened while the system is powered, this could potentially result in an arc flash that could lead to serious injury to service personnel in immediate proximity.

The problem would only present itself during servicing activities involving the electronics cabinet. The electronics cabinet is located within the equipment room and should only be accessed by qualified service personnel.

To date, there have been no injuries reported to GE HealthCare as a result of this problem.
Instructions for Customers/Users Pending correction by GE HealthCare, customers can continue to use the system.

The electronics cabinet door should be closed any time there is power to the system. In the event the electronics cabinet needs to be serviced, customers are to contact their local GE HealthCare service representative or authorised third-party service providers.

A GE HealthCare representative will contact customers to arrange a site visit to undertake the correction.
Action Commencement Date 22/10/2026
Responsible Entity GE Healthcare Australia Pty Ltd
Contact Information 1800 659 465 - GE HealthCare Service