| TGA Action ID |
RC-2026-RN-00812-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Eye-Light, My Mask-E and Meibomask
All serial numbers |
| ARTG Number |
328245 |
| Action Type |
Product Correction |
| Action Level |
Wholesale |
| Hazard Classification |
Class II |
| Reason for the Action |
Investigation found that use of the EYE-MASK LED mask in patients with certain anatomical or ophthalmic conditions may reduce the required 5–10 mm distance between the skin and the LEDs, resulting in excessive localised heating. These conditions include exophthalmos or proptosis, periocular oedema and dermatochalasis.
The manufacturer is updating the user manuals for EYE-LIGHT, MY MASK-E and MEIBOMASK to include additional warnings for use in patients with these conditions. |
| Instructions for Customers/Users |
Customers are to: - Replace the affected user manuals with the latest revisions provided by the manufacturer. - Ensure relevant staff are informed of the new warnings and precautions and use the updated manuals. |
| Action Commencement Date |
3/11/2026 |