| TGA Action ID |
RC-2026-RN-00804-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Philips Radiography 7000 M
Product Number: 712010
Affected Software: R1.0, R1.0.1, R1.0.2, and R1.0.3 |
| ARTG Number |
451166 |
| Unique Device Identifier (UDI) |
N/A |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class III |
| Reason for the Action |
When the operator is switching between examinations, an intermittent software failure may cause the console to display the message "Call service! Filter selection failure." This display message may appear even if the motorised filter has been selected correctly. If this happens, the system enters a Not Ready state and does not permit an X-ray exposure. |
| Instructions for Customers/Users |
Customers are to: - Continue to use the affected system as intended. - If the filter selection failure message appears, press OK to acknowledge it. If the message clears and the system returns to Ready, continue use. If not, reboot the system and confirm the system returns to Ready before proceeding. - If the system does not return to Ready after rebooting, or the message recurs, contact Philips Service.
Philips will contact customers to schedule a time to implement the software solution beginning Q4 2026. |
| Action Commencement Date |
13/10/2026 |
| Responsible Entity |
|
| Contact Information |
1800 251 400 - Philips Service Delivery Team |