TGA Action ID RC-2026-RN-00800-1
Type of Product Medical Device
Product Name/Description Philips MR systems

Ingenia Ambition S
Model Numbers: 781359, 782108, 782133, 782139

Ingenia Ambition X
Model Numbers: 781356, 782109, 782138

Ingenia Elition S
Model Numbers: 781357, 782106, 782137, 782150

Ingenia Elition X
Model Numbers: 781358, 782107, 782119, 782136, 782151

MR 7700
Model Numbers: 782120, 782153, 782129

Multiple Serial Numbers
ARTG Number 230064, 372370, 485786
Unique Device Identifier (UDI) N/A
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Philips has identified a problem with MR system patient supports manufactured between June 2022 and January 2024. Over time, vertical movement of the MR table can gradually damage cables connected to the vertical drive unit capacitor, which could result in an electrical short circuit.

If a short circuit occurs, the electrical mains breaker trips as designed in accordance with IEC 60601-1 requirements, causing loss of system power and MR system downtime. This would result in a system shutdown and temporary unavailability of the MR system until service is performed.

No patient injury or other adverse clinical outcome has been reported to Philips in relation to this problem.
Instructions for Customers/Users If a short circuit occurs, customers are to contact their Philips representative.

Philips will contact customers to schedule a site visit to install a hardware correction (reference FCO78100622).
Action Commencement Date 13/10/2026
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team