Print version of this report
TGA Action ID RC-2026-RN-00787-1
Type of Product Medical Device
Product Name/Description Sevoflurane Vaporizers

Vaporizer Sevoflurane Safe-T-Seal
Part Number: 6887135
Serial Numbers: 30144 - 30731, 31026 - 31061, 31066, 31069, 31070

Vaporizer Sevoflurane Maquet filling
Part Number: 6682282
Serial Numbers: 30000 - 30911, 33065, 33074, 33077, 33078, 33086
ARTG Number 180464
Action Type Recall
Action Level Hospital
Hazard Classification Class I
Reason for the Action Maquet Critical Care AB has identified a problem affecting certain batches of Sevoflurane vaporizers that were not manufactured according to specification. Components within the affected vaporizers may not perform as intended over time, which could affect the long-term performance of the devices.

To date, no patient or operator injuries have been reported. Affected vaporizers are being recalled and replaced with conforming units.
Instructions for Customers/Users Customers are to:
- Continue using affected vaporizers in accordance with the instructions provided in the customer letter until replacement vaporizers are supplied.
- Empty and dry-run the vaporizer if sevoflurane is not expected to be used within the next 30 days or before external transportation.
- Before use, verify that sevoflurane has been used within the previous 30 days; if not, discard the sevoflurane and empty and dry-run the vaporizer in accordance with the provided instructions.
Action Commencement Date 15/10/2026
Responsible Entity Getinge Australia Pty Ltd
Contact Information 1800 438 464 - Customer Service