TGA Action ID RC-2026-RN-00774-1
Type of Product Medical Device
Product Name/Description HAMILTON-C6 Ventilator

Product Code: HMT-160021

Part Number: 160021

Software Version: v2.0.1
ARTG Number 113999
Unique Device Identifier (UDI) 07630002808590
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Hamilton Medical has identified several software problems that may occur under specific operating conditions in devices running software version 2.0.1. These problems may result in false disconnection alarms, temporary panel reboots, unsuccessful flow sensor calibration, or the ventilator continuing to apply a previous pressure limit (Plimit) setting after certain ventilation mode changes.

To date, no reports of patient harm, serious injury, or death related to these problems have been received.
Instructions for Customers/Users Customers are to identify all devices operating with software version 2.0.1 within their facility and continue using affected devices in accordance with the precautions outlined in the customer letter while awaiting installation of software version 2.0.2.
Action Commencement Date 2/10/2026
Responsible Entity Device Technologies Australia Pty Ltd