TGA Action ID RC-2026-RN-00759-1
Type of Product Medical Device
Product Name/Description ARTIS pheno, ARTIS icono floor, ARTIS icono biplane, ARTIS icono ceiling with software version VE50B (only systems that were upgraded to VE50B)

ARTIS pheno
Product Code: 10849000

ARTIS icono floor
Product Code: 11327700

ARTIS icono biplane
Product Code: 11327600

ARTIS icono ceiling
Product Code: 11328100
ARTG Number 273952
Unique Device Identifier (UDI) 04056869063317, 04056869295923, 04056869149325, 04056869046877
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action In certain operating situations, the system may fail to execute movement control commands as intended. If this occurs, motorised stand and/or table movement may no longer be available until the system has been completely restarted. The impact on system operation is that movement of the affected system axes may no longer be possible until the system is restarted.

This may limit or delay the ability to change projection planes or to position the system as required for the intended procedure. In situations where immediate access to the patient or repositioning of the system is required, including during cardiopulmonary resuscitation (CPR) activities, the described behaviour may interfere with the intended workflow.
Instructions for Customers/Users Customers should read the customer letter and perform all instructions provided in this letter.

Ensure that all users of the affected system are informed of this problem and are familiar with the applicable patient rescue and emergency procedures described in the Operator Manual.

No specific retrospective patient follow-up is currently considered necessary with regard to this problem.

The manufacturer is currently investigating the underlying cause of the reported system behaviour with high priority. Until the root-cause investigation has been completed no additional specific risk mitigation measures beyond the information provided in this letter have been defined at this time.

If additional actions are identified as necessary based on the outcome of the investigation, affected customers will be informed accordingly.

UPDATE 23/10/26:
Siemens Healthineers has released corrective software update Patch 3.5 which provides a recovery function if the problem occurs.

A Siemens Healthineers service organisation will contact customers to arrange an appointment to install the applicable corrective software update on affected systems.

This action has been closed out on 05/11/2026
Action Commencement Date 30/09/2026
Responsible Entity Siemens Healthcare Pty Ltd
Contact Information 1800 310 300 - Siemens Customer Care Centre