TGA Action ID RC-2026-RN-00758-1
Type of Product Medical Device
Product Name/Description HemosIL AcuStar HIT-IgG(PF4-H). An in vitro diagnostic medical device (IVD).

Part Number: 0009802028

Lot Number: B38063

Expiry Date: 01/11/2026
ARTG Number 221296
Unique Device Identifier (UDI) 08426950545774
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Werfen has identified a performance problem affecting Lot B38063 of HemosIL AcuStar HIT-IgG (PF4-H). The problem may cause negative bias in test results, resulting in some samples being reported below the assay cut-off of 1.00 U/mL. As a result, HIT IgG antibody-positive samples may be incorrectly reported as negative.
Instructions for Customers/Users Customers are to:
- Destroy Lot B38063 in accordance with their standard procedures
- Contact Werfen Australia for alternative testing options
Action Commencement Date 29/09/2026