TGA Action ID RC-2026-RN-00745-1
Type of Product Medical Device
Product Name/Description Ventilation (tympanostomy) Tube 5PK Button 1.27 FLPL

Catalogue Number: 1013301

Lot Number: 0231048280

GTIN: 20763000281636
ARTG Number 126695
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Medtronic is recalling a certain lot of Ventilation (tympanostomy) Tubes that were distributed without undergoing the required sterilisation process.

After implantation this could cause infection and subsequent inflammation potentially leading to temporary hearing impairment, otorrhea (fluid discharge) and the need for antibiotics.

As of September 10, 2026, Medtronic has received no complaints associated with the affected lot and no reports of patient harm related to these devices.
Instructions for Customers/Users Customers are to:
- Stop using and quarantine the affected product.
- Return unused affected stock following the instructions in the customer letter.

Medtronic advises that explant is not recommended for devices already implanted as the devices are fully functional and continue to provide their clinical benefit.

As a precautionary measure, patients implanted with devices from affected lot numbers should continue to be monitored in accordance with standard clinical practice.
Action Commencement Date 25/09/2026
Responsible Entity Medtronic Australasia Pty Ltd