Print version of this report
TGA Action ID RC-2026-RN-00739-1
Type of Product Medical Device
Product Name/Description KayserBetten Safe Surround Genie Bed

Product code: 7959-0016-000

Date of Manufacture: Before or on 9 July 2025

Multiple serial numbers
ARTG Number 151342
Action Type Product Correction
Action Level Consumer
Hazard Classification Class I
Reason for the Action The KayserBetten Safe Surround Genie Bed hand control does not automatically lock after use. If it is not locked with the key when the user is unattended, the user or another person may operate the bed. Someone could mistakenly use the hand control and accidentally move the bed.

This could cause trapping beneath or between moving parts, resulting in serious injury or death. Two (2) deaths have been reported in relation to this problem.
Instructions for Customers/Users Customers are to:
Strictly observe the following interim safety measures, until the new Auto-Lock hand control is fitted.

Whenever the user is unattended:
· Move the sleeping platform to a level position and its lowest height.
· Lock all hand control functions using the key
· Close the bed doors

Medifab will contact customers to organise an appointment. At this appointment a Medifab staff member will remove the original hand control, install the new Auto-Lock hand control, perform a functional check, and explain its operation to the user or caregiver.
Action Commencement Date 1/10/2026
Responsible Entity Medifab Ltd
Contact Information 1300 543 343 - Medifab Australia