TGA Action ID RC-2026-RN-00720-1
Type of Product Medical Device
Product Name/Description BD Alaris Point of Care Unit (PCU) Wireless ONLY

Model Number: 8015

Software Version: 12.1.1
ARTG Number 146664
Unique Device Identifier (UDI) 10885403515231, 10885403516023, 10885403519215, 10885403519284, 10885403519291
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Device labelling does not indicate that Transport Layer Security (TLS) versions must be enabled on the hospital network to ensure secure Wi-Fi communication with the Alaris and BD Alaris Point-of-Care Unit (PCU) when using enterprise-level Wi-Fi security.

Without clear labelling, hospitals may unknowingly disable TLS 1.0, resulting in network connectivity failure and memory leaks which could lead to potential System Error.
Instructions for Customers/Users BD is updating its product labelling for actively installed versions of Systems Manager to clarify compatibility requirements related to TLS protocols to ensure customers can appropriately configure their networks.
Action Commencement Date 23/09/2026
Responsible Entity Becton Dickinson Pty Ltd
Contact Information (02) 8875 7000 - BD